Linac Daily Output QA Devices : Sun Nuclear SRS PROFILER

Sun Nuclear SRS PROFILER

Sun Nuclear SRS Profiler

Redefining Radiosurgery QA

Sophisticated beam QA for stereotactic radiosurgery fields measuring output, flatness, symmetry and field size with a single exposure.

The SRS Profiler Advantage
  • The first QA device designed for cone based SRS
  • Quick to setup and easy to use
  • Attaches to the Accuray® Cyberknife® birdcage for measurements at any angle
  • Streamlines acceptance testing, commissioning, and routine QA tests
  • Operation using PROFILER analysis software
Specifications
  • Detector type: Sun Nuclear Diode Detectors
  • Detector quantity: 125 total on four axes
  • Axes length (mm): 143 (X), 127 (Y), 124 (Diagonals)
  • Detector spacing (mm): 4.0, 2.0 with radial offset
  • Radial field size (mm): 10, 20, 40, 60, 80, 100 and 120
  • Active detector area (mm2): 0.64
  • Detector volume (cm3): 0.000019
  • Detector sensitivity (nC/Gy): 32.0
  • Detector stability: 0.5%/kGy at 6MV
  • Dose rate dependence: ±1%, 75 - 250cm SSD
  • Energies: >1MV, >=4 MeV
  • Dose rate limit (gy/min): > 20
  • Sampling rate (ms): 50
  • Operating system: Windows 2000, XP 32-bit, or Vista 32-bit
  • Dimensions (cm)/ Weight (kg): 4.3 x 15.0 x 33.5 / 1.5
  • Number of connection cables: Single power/data cable

Click Here to Download a Printable Information Sheet. (.PDF format)